Appendix B: Performance Verification
BĆ28
DG2040 Data Generator
5. Change the oscilloscope horizontal axis setting to 2.00 s/div.
6. Verify that the sync signal appears in every 10 s on the oscilloscope screen.
7. Press the following buttons to set the DG2040 Data Generator trigger
interval to 10 ms.
Menubutton
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
Interval
State (Set to On)
Time
10
kHz/ms/mV
8. Change the oscilloscope horizontal axis setting to 2.00 ms/div.
9. Press the following buttons to set the DG2040 Data Generator trigger
interval to turn off.
Menubutton
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
Interval
State (Set to Off)
10. Verify that Waiting for trigger message appears on the DG2040 Data
Generator screen, and no waveform appears on the oscilloscope screen.
11. Press the following buttons to set the DG2040 Data Generator trigger
interval to Ext.
Menubutton
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
SourceĂ
(set to Ext)
12. Verify that the sync signal appears every 2 ms on the oscilloscope screen.
13. Press the START/STOP button on the front panel to stop output. The LED,
corresponding to the button, is turned off.