![Tektronix 59021-10 User Manual Download Page 224](http://html1.mh-extra.com/html/tektronix/59021-10/59021-10_user-manual_1078414224.webp)
Appendix B: Performance Verification
DG2040 Data Generator
BĆ31
H
Function generator
Output channel
CH2
. . .
Waveform
Square wave
. . . . . . .
Frequency
10 MHz
. . . . . . .
Amplitude
2 V
. . . . . . .
Offset
0 mV
. . . . . . . . . .
Procedure.
Follow the steps below to verify that the edge control mode operation
is functioning correctly.
1. Load the TP7EDGE.PDA test pattern file from the performance check disk.
2. Press the START/STOP button on the front panel. The LED, corresponding
to the button, should be on.
3. Adjust the Main Pos control on the sampling oscilloscope front panel. Verify
that jitter appears only on the rising edge portion on the CH0 output signal
after the CH1 output signal rises. See figure B-16.
4. Press the START/STOP button on the front panel to stop output. The LED,
corresponding to the button, is turned off.
CH1
(CH0 OUT)
CH2
(CH1 OUT)
Figure BĆ16: Edge control mode
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com