Appendix B: Performance Verification
BĆ10
DG2040 Data Generator
Table BĆ4: Internal clock frequency accuracy
Internal clock frequencies
Clock frequency ranges
1.100000 GHz
1.0999989 GHz
1.1000011 GHz
650.1000 MHz
650.09935 MHz
650.10065 MHz
200.0000 kHz
199.99980 kHz
200.00020 kHz
100.0000 mHz (10.0 s)
99.99990 mHz
100.00010 mHz
6. Press the START/STOP button on the front panel to stop the output (the
buttons LED goes off).
This test verifies that the pattern data that is output from the DG2040 Data
Generator output channel ports (CH0, CH0, CH1, and CH1) is functioning
correctly.
The following equipment is required.
Equipment Required
Oscilloscope
Function generator
Three BNC cables
Two SMA cables
Two SMA femaleĆtoĆBNC male adapters
Performance check disk
Connections.
Refer to Figure B-4 for connections.
Use the SMA cables and the SMA female-to-BNC male adapters to make the
following connections:
H
Connect CH0 output from the DG2040 Data Generator front panel to the
oscilloscope CH1 input.
H
Connect CH0 output from the DG2040 Data Generator front panel to the
oscilloscope CH2 input.
Use the BNC cables to make the following connections:
H
Connect the function generator CH1 output to the DG2040 Data Generator
rear panel event input.
H
Connect the event output from the DG2040 Data Generator rear panel to the
oscilloscope CH3 input.
Sequence & Data Output