7-14
920017-148
Operator Manual
Troubleshooting
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout (Cont'd)
PROBLEM
POSSIBLE CAUSE AND/OR CORRECTION
32. Alarm sounds and screen displays:
ALARM: ACID
PUMP IS DEFECTIVE
Printer records:
* ALARM
ACID
PUMP IS DEFECTIVE
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Acid chemical injection pump still energized even if control de-
energized pump — Call STERIS.*
33.
If pH Neutralizing System option is
present,
alarm sounds and screen dis-
plays:
ALARM: pH INJECTION
TIME TOO LONG
Printer records:
* ALARM
pH INJECTION
TIME TOO LONG
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Chemical container empty. — Replace empty container.
3. pH injection time longer than 15 minutes — Call STERIS.*
*Service charges may be incurred. Consult your warranty for details.