4-7
Operating Instructions
Operator Manual
920017-148
DOOR
OPEN
2.
Wait for chamber doors to open and push loaded accessory cart into wash
chamber.
-- or --
If photoelectric sensor is disabled
, press
DOOR OPEN
touch pad to open
door and push cart into wash chamber. For procedure on how to disable
photoelectric sensor in Service Mode, see Section 4 in
Maintenance Manual
(P764330-842).
IMPORTANT:
If the Central Header Manifold is used, assure Floor Tilting option
IS NOT ACTIVATED in the wash chamber where it is positioned (see Section 5,
Cycle and Control Value Programming).
1.
When doors are open, enter wash chamber and roll Central Spray Header
Manifold and Support (see Figure 4-2) along central rails until top of header
is aligned with roof locking support.
2.
Unlock safety locking pin retaining Central Header Manifold to support.
3.
Align manifold at the bottom of the accessory with the water inlet connector
at the bottom of the wash chamber.
4.
Push up locking handle towards roof locking support and turn clockwise to
secure header into position.
5.
Roll Support out from wash chamber.
6.
Roll Rodent Cage Racks (see Figure 4-3) along Central Spray Header
Manifold inside wash chamber.
4.4.3 How to Load
Central Header Manifold
with Rodent Cage Racks
(Accessory)
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
•
Always position mani-
folded Bottle Washing Cart
and/or Central Header
Manifold over central wa-
ter inlet connector. If mani-
folded accessory is not
positioned correctly, dam-
age may result and unit
will be unable to effectively
wash load.
•
Do not process load using
Central Header Manifold or
Bottle Washing Cart in a
chamber where Automatic
Floor Tilting option is acti-
vated. If Automatic Floor
Tilting is used, manifolded
water inlet and washer will
be damaged.