7-10
920017-148
Operator Manual
Troubleshooting
19. Load
DOOR OPEN touch pad was pressed
with load-end door open. Screen displays:
ALARM:LOAD DOORS
ARE OPEN
Printer records:
* ALARM
LOAD DOORS
ARE OPEN
HH:MM:SSX
YY/MM/DD
Table 7-2. Troubleshooting Chart – Alarm Screen and/or Printout (Cont'd)
PROBLEM
POSSIBLE CAUSE AND/OR CORRECTION
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Close load doors. Press CYCLE START to resume.
3. Loss of air supply during a cycle — Call STERIS.*
4. Door sensors defective — Call STERIS.*
5. If situation reoccurs, call STERIS.*
More than 20 seconds needed to close unload doors.
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Unload doors are obstructed — Remove obstruction behind or
between doors.
3. Air supply valves closed – Open air supply valves.
4. Air pressure is below specifications – Call STERIS.*
5. If condition reoccurs, call STERIS.*
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Unload doors are obstructed — Remove obstruction behind or
between doors.
3. Air supply valves closed – Open air supply valves.
4. Air pressure is below specifications – Call STERIS.*
5. If condition reoccurs, call STERIS.*
17. Unload door remains closed after DOOR
OPEN touch pad was pressed. Alarm
sounds and screen displays:
ALARM: UNLOAD DOORS
DID NOT CLOSE
Printer records:
* ALARM
UNLOAD DOORS
DID NOT OPEN
HH:MM:SSX
YY/MM/DD
18. Unload door remains open after DOOR
CLOSE touch pad was pressed. Alarm
sounds and screen displays:
ALARM: UNLOAD DOORS
DID NOT CLOSE
Printer records:
* ALARM
UNLOAD DOORS
DID NOT CLOSE
HH:MM:SSX
YY/MM/DD
*Service charges may be incurred. Consult your warranty for details.