6-15
Routine Maintenance
Operator Manual
920017-148
If Automatic Descaler System option is present
, an additional chemical pump
and flowmeter allows operator to process a DESCALER cycle during which
Liquid Descaler Liquid Scale Remover is automatically injected into load and
unload chambers.
NOTE: When processing a DESCALER Cycle, note the following:
1) Automatic DESCALER Cycle has a duration of one hour.
2) If unit has been configured with Central Spray Header Manifold or Bottle
Washing Cart options, you must use accessory to perform DESCALER Cycle.
1.
If Central Header Manifold or Bottle Washing Cart is present
, roll Central
Spray Header Manifold or Bottle Washing Cart into wash chamber.
2.
Connect manifold coupling system (option) to accessory (see Section 4.4,
Accessories).
3.
Press
CYCLE MENU
touch pad to access third Factory-Set cycle menu.
Display shows:
CYCLE 9
CYCLE 10
DRAIN
DESCALER
___ indicates flashing position
4.
Press
SELECT CYCLE
touch pad to select DESCALER (flashing). Press
CYCLE START
touch pad to confirm selection and start DESCALER Cycle.
Display shows:
DESCALER
ALKALINE
DRAINING
... then:
DESCALER
RINSE
DRAINING
... then:
DESCALER
RINSE
COOLDOWN
...then:
DESCALER
ACID
DRAINING
5.
All tanks will drain completely to assure only clean water is used.
6.3.2 Automatic
DESCALER Cycle (Option)
C A U T I O N – P O S S I B L E
EQUIPMENT DAMAGE: Do
not process load using Cen-
tral Spray Header Manifold,
Bottle Washing Cart, or Uni-
versal Cage and Pan Wash
Cart in a chamber for which
Floor Tilt option is activated.
If Floor Tilt is used with these
accessories, manifolded wa-
ter inlet and washer will be
damaged.
CYCLE
MENU
SELECT
CYCLE
CYCLE
START
Name of Cycle
Started
Phase in
Progress
Treatment in
Progress