2-1
Installation Verification
Operator Manual
920017-148
INSTALLATION VERIFICATION
IMPORTANT:
A listing of the
Safety Precautions to be observed when operating
and servicing this equipment can be found in Section 1 of this manual. Do not
operate or service the equipment until you have become familiar with this
information.
These specifications are intended to describe the technical information given
on the nameplate of your washer and to state other relevant information. Check
Equipment Drawing or Identification nameplate, located on frame of mobile
mechanical core, below main electrical box, for proper voltage and amperage.
This unit operates on:
• 380/400/415 V~, 3-phase, 50 Hz
• 480 V~, 3-phase, 60 Hz.
• 208 V~, 3-phase, 60 Hz (with the optional transformer).
A protective ground conductor is required (Class 1 Equipment). Installation
Category is Overvoltage II.
Refer to
Uncrating/Installation Manual (P920017-147) for proper connection.
IMPORTANT:
Customer responsible for compliance with applicable codes and
regulations.
The maximum currents and power consumptions, for units with or without
drying, are:
kW
A
• 380/400/415 V
18,6
26
• 480 V
28,3
34
Equivalent Sound Pressure Level at work station (measured 3 ft. [1 m] away from
equipment and at 5 ft [1.6 m] from ground):
76.6 dB (A)
for a freestanding unit
with side panels (Results determined according to
ISO-3746: 1979 Standard:
Acoustics Determination of Sound Power Levels of Noise Sources Survey
Method).
This washer is designed to give optimal results in an environment where
maximum relative humidity is less than 85% and maximum operating tempera-
ture is 104°F (40°C).
2.1.1 Voltage and
Amperage
2
2.1 Technical
Specifications
2.1.2 Noise Level
2.1.3 Permissible
Environmental Conditions