4-13
Operating Instructions
Operator Manual
920017-148
DRAIN cycle can be performed whenever necessary, after each cycle, between
different load types, or at the end of a work session.
1.
Press
CYCLE MENU
touch pad to scroll to appropriate cycle menu. Display
shows:
CYCLE 9
CYCLE 10
DRAIN
DESCALER
___ indicates flashing position
2.
Press
SELECT CYCLE
touch pad to select DRAIN Cycle. Display shows:
CYCLE 9
CYCLE 10
DRAIN
DESCALER
___ indicates flashing position
3.
Press
CYCLE/START
touch pad to confirm. Display shows:
PURGE TRAVELERS?
CVTP=YES
SVTP=NO
NOTE: Purging travelers will also purge Rinse Tank (Tank 2).
4.
Press
CHANGE VALUES
touch pad to purge travelers or press
SAVE
VALUES
touch pad to skip to the next question. Display shows:
DRAIN ALL TANKS?
CVTP=YES
SVTP=NO
5.
Press
CHANGE VALUES
touch pad to select YES and drain all tanks. Control
will automatically move to
Step 9
.
-- or --
Press
SAVE VALUES
touch pad to select NO and choose which tanks to
drain. Display shows:
DRAIN ALKALINE
CVTP=YES
SVTP=NO
6. Press
CHANGE VALUES
touch pad to drain Alkaline Tank (Tank 1), or press
SAVE VALUES
touch pad to keep tank filled. Display shows:
DRAIN RINSE?
CVTP=YES
SVTP=NO
NOTE: The display is only shown if
NO
has been selected in answer to
PURGE
TRAVELERS?
7. Press
CHANGE VALUES
touch pad to drain Rinse Tank (Tank 2), or press
SAVE VALUES
touch pad to keep tank filled. Display shows:
DRAIN ACID?
CVTP=YES
SVTP=NO
NOTE: The display appears only if Additional Wash Tank option is present.
4.6 DRAIN Cycle
CYCLE
START
SAVE
VALUES
CHANGE
VALUES
SELECT
CYCLE
CHANGE
VALUES
SAVE
VALUES
SAVE
VALUES
CHANGE
VALUES
SAVE
VALUES
CHANGE
VALUES
CYCLE
MENU