7-5
Troubleshooting
Operator Manual
920017-148
Table 7-1. Troubleshooting Chart – No Alarm Screen or Printout (Cont'd)
PROBLEM
POSSIBLE CAUSE AND/OR CORRECTION
Call STERIS.*
1. Cable and pulley malfunction – Call STERIS.*
2. Door rails malfunction in wash chamber – Call STERIS.*
3. If condition reoccurs, call STERIS.*
1. Doors obstructed – Remove obstruction.
2. Cables and pulleys malfunction – Call STERIS.*
3. Doors guiding system malfunction – Call STERIS.*
4. Floor tilt still in upper position – Call STERIS.*
5. Air flow control improperly adjusted – Call STERIS.*
6. Air pressure improperly adjusted – Call STERIS.*
1. Air flow control improperly adjusted – Call STERIS.*
2. Air pressure improperly adjusted – Call STERIS.*
1. Cable(s) and pulley(s) malfunction – Call STERIS.*
2. Air flow control improperly adjusted – Call STERIS.*
3. Air pressure improperly adjusted – Call STERIS.*
1. Cable and pulley malfunction – Call STERIS.*
2. Air flow control improperly adjusted – Call STERIS.*
3. Air pressure improperly adjusted – Call STERIS.*
1. Load exceeds 800 lb (363 kg) – Check load weight.
2. If condition reoccurs, call STERIS.*
1. Insufficient time set for drying phase – Increase drying time.
2. If condition reoccurs, call STERIS.*
1. Self-cleaning filter clogged – Clean filter.
2. If condition reoccurs, call STERIS.*
1. Fuse is blown – Call STERIS.*
2. Defective relay – Call STERIS.*
3. Defective chemical pump – Call STERIS.*
28. Suction pump abnormally noisy.
29. Doors open and /or close too slowly and/
or vibrate.
30. Doors do not open completely.
31. Doors open and/or close too quickly.
32. Travelers vibrating.
33. Travelers are moving too slowly.
34. Auto-tilt floor lifting does not operate prop-
erly.
35. Incomplete Drying.
36. Abnormal draining of wash chamber.
37. Chemical pump(s) malfunction.
*Service charges may be incurred. Consult your warranty for details.