7-6
920017-148
Operator Manual
Troubleshooting
Table 7-2. Troubleshooting Chart - Alarm Screen and/or Printout
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Supply valves closed – Open supply valves.
3. Building air supply too low – Call STERIS.*
4. STOP/RESET touch pad is pressed during the opening of the
Unload Door. If operator does not press CYCLE START within
4 minutes, alarm will reoccur every 4 minutes (default) until a
new cycle is started.
1. Tank too long to fill. Alarm sounds and screen
displays:
ALARM:
(TANK NAME)
TOO LONG TO FILL
Printer records:
* ALARM
(TANK NAME)
TOO LONG TO FILL
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Detergent empty – Replace container.
3. If condition reoccurs, call STERIS.*
3. Alarm sounds and screen displays:
ALARM:
(DET. NAME) INJ
NOT REACHED
Printer records:
* ALARM
(DETERGENT NAME) INJ
NOT REACHED
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Debris screen filter clogged – Clean debris screen.
3. Self-cleaning filter clogged – Clean filter.
4. If condition reoccurs, call STERIS.*
2. Tank too long to drain. Alarm sounds and
screen displays:
ALARM:
(TANK NAME)
TOO LONG TO DRAIN
Printer records:
* ALARM
(TANK NAME)
TOO LONG TO DRAIN
HH:MM:SSX
YY/MM/DD
1. Press ALARM REPLY touch pad to silence alarm buzzer.
2. Supply valves closed – Open supply valves.
3. Building air supply too low – Call STERIS.*
PROBLEM
POSSIBLE CAUSE AND/OR CORRECTION
4. Alarm sounds and screen displays:
ALARM: LOAD PUMP
PRESSURE IS TOO LOW
Printer records:
* ALARM
LOAD PUMP
PRESSURE TOO LOW
HH:MM:SSX
YY/MM/DD
*Service charges may be incurred. Consult your warranty for details.