5-9
Cycle and Control Value Programming
Operator Manual
920017-148
39. To enter DRYING treatment temperature, press
VALUE
(up or down) touch
pad to select temperature. Treatment temperature is input in Fahrenheit or
in Celsius, according to Service Mode setting (see Section 4 of Mainte-
nance Manual [P764330-842]), within a range of 140 to 180°F (60-82.2°C).
40. Press
CHANGE VALUES
touch pad to confirm temperature selection.
41. The REFRESH RATE feature allows operator to refresh chemical solutions
with fresh water after each cycle. REFRESH RATE is adjustable within a
range of 15 to 100%. If the REFRESH RATE is set at 100%, the chemical
tank will be completely drained and filled with fresh water.
The ALKALINE tank REFRESH RATE screen appears. Display shows:
ALKALINE REFRESH
15%
__indicates flashing position.
To enter the ALKALINE tank REFRESH RATE, press
VALUE
(up or down)
touch pad to select percentage. REFRESH RATE can be adjusted within
a range of 15 to 100%.
42. Press
CHANGE VALUES
touch pad to confirm REFRESH RATE selection.
If unit is equipped with a second chemical tank, display shows:
ACID REFRESH
15%
__indicates flashing position.
43 To enter the ACID Tank REFRESH RATE value, press
VALUE
(up or down)
to select percentage. REFRESH RATE can be adjusted within a range of
15 to 100%.
44. Press
CHANGE VALUES
touch pad to confirm REFRESH RATE.
45.
If unit is equipped with Automatic Floor Tilting option
, the load
AUTOMATIC FLOOR TILTING screen appears. Display shows:
LOAD FLOOR TILT ?
NO
__indicates flashing position.
IMPORTANT:
This screen is to activate or deactivate automatic floor tilting
for the load chamber only.
Press
VALUE
(up or down) touch pad to change YES to NO. Press
CHANGE
VALUES
touch pad to confirm selection.
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼
CHANGE
VALUES
CHANGE
VALUES
CHANGE
VALUES
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: Do
not process load using Cen-
tral Header Manifold or
Bottle Washing Cart in a
chamber where Automatic
Floor Tilting option is acti-
vated. If Automatic Floor
Tilting is used, manifolded
water inlet and washer will
be damaged.
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼
CHANGE
VALUES