5-7
Cycle and Control Value Programming
Operator Manual
920017-148
21. Press
VALUE
(up or down) touch pad to select between YES or NO.
NOTE: MIDRINSE is non-recirculated and time setting is not adjustable.
22. Press
CHANGE VALUE
touch pad to confirm selection. The WASH 2 time
value screen appears. Display shows:
WASH 2
00:00
__indicates flashing position.
23. To enter WASH 2 treatment time, press
CURSOR
(left or right) touch pad
and
VALUE
(up or down) touch pad. Treatment time is input in minutes and
seconds within a range of 0 to 10 minutes.
24. Press
CHANGE VALUES
touch pad to confirm selection.
NOTE: WASH 2 sequence continues only if time selected is different from 0.
25. If unit is equipped with an Additional Wash Tank, the WASH 2 tank name
screen appears. Display shows:
WASH 2 USES
ACID
__indicates flashing position.
NOTE: If unit is not equipped with an Additional Wash Tank, control will move
on to the WASH 2 chemical rate message.
26. Press
VALUE
(up or down) touch pad to change ACID to ALKALINE. Press
CHANGE VALUES
touch pad to confirm selection. The WASH 2 chemical
rate value screen appears. Display shows, for example:
SET ACID RATE
0 OZ/GAL
__indicates flashing position.
27. Press
VALUE
(up or down) touch pad to select chemical rate. Chemical rate
is input in oz/gal or in mL/L, according to Service Mode setting (see
Section 4 of
Maintenance Manual [P764330-842]), within a range of 0 to
4 oz/gal (0-32mL/L) in 1/8 oz (2 mL) increments.
28. Press
CHANGE VALUES
touch pad to confirm chemical entry. The
WASH 2 temperature value screen appears. Display shows:
WASH 2
160F
__indicates flashing position.
29
. Press
VALUE
(up or down) touch pad to select WASH 2 temperature.
Temperature is input in Fahrenheit or in Celsius, according to Service Mode
setting (see Section 4 of
Maintenance Manual [P764330-842]), within a
range of 120 to 165°F (49 to 74°C).
CHANGE
VALUES
CHANGE
VALUES
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼
▲▲▲▲▲
▼▼▼▼▼
CURSOR
CHANGE
VALUES
CHANGE
VALUES
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼