3-9
Component Identification
Operator Manual
920017-148
Door panels can be pushed out in case of an emergency, offering four
emergency exit points (see Figure 3-11).
Four EMERGENCY STOP cables, located on each side of wash chamber,
instantly stop washer operation when pressed (see Figure 3-11).
Two EMERGENCY STOP pushbuttons located under each control panel,
instantly stop washer operation when pressed.
NOTE: EMERGENCY STOP pushbuttons will cut off electrical power to control
only and completely depressurize unit.
Interior light flashes once prior to cycle start, warning the operator that a cycle
is about to begin.
An adjustable Safety Delay (15 to 60 seconds), during which the washer will
sound an alarm signal, has been provided to warn operator to leave the wash
chamber before the cycle starts. Safety Delay is adjustable in Service Mode
(see Section 4 of
Maintenance Manual [P764330-842]). Default Safety Delay is
15 seconds.
3.5.2 Emergency Exit
Safety Doors
3.5.4 EMERGENCY STOP
Pushbuttons
3.5.3 EMERGENCY STOP
Cables
WARNING – PERSONAL
INJURY AND/OR EQUIP-
MENT DAMAGE HAZARD:
•
To open doors from inside
wash chamber, pull EMER-
GENCY STOP cables.
Washer operation will auto-
matically stop. Then, push
firmly between door panels
using shoulder and upper
arm, applying upper body
force. Do not push between
the two doors, but between
the door panels.
•
To test or demonstrate
Emergency Exit Safety
Doors, first press EMER-
GENCY STOP pushbutton
(located under control) or
EMERGENCY STOP cables
(inside wash chamber) to
turn power OFF. If power is
still on while adjusting or
servicing doors, the photo-
electric sensor will detect
the movement of the door
panels and doors will open
automatically.
3.5.5 Interior Light
3.5.6 Safety Delay