4-2
920017-148
Operator Manual
Operating Instructions
The system is self-priming and requires no specific operation. At the beginning
of the cycle, display shows for example:
PRIMATES STARTING
PRIMING
If chemical flowmeters detect no detergent is being injected, an alarm buzzer
sounds and display shows:
ALARM: (CHEM NAME) INJ
NOT REACHED
Chemical Container Replacement, for procedure on how to
replace chemical containers.
To achieve maximum cleaning efficiency, select chemical appropriate to soil
type being processed. See Section 2.3,
Chemical Additives Specifications.
4.2 Priming
Procedure
WARNING – CHEMICAL
BURN AND/OR EYE INJURY
HAZARD:
•
Always use non-foaming
chemical for effective clean-
ing and proper pump and
water level control opera-
tion. Follow manufacturer's
recommendations for amount
of chemical to be used.
•
Detergents are caustic and
can cause adverse effects to
exposed tissues. Do not get
in eyes, on skin or attempt
to swallow. Read and follow
the precautions and instruc-
tions on the detergent label
and in the Material Safety
Data Sheet (MSDS) prior to
handling detergent contain-
ers, or servicing the deter-
gent injection pumps and
lines. Wear appropriate per-
sonal protective equipment
(PPE) whenever handling
chemicals or servicing
chemical injection pumps
and lines.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
When choosing a chemical,
select one with a low chlo-
ride content. Chloride is
harmful to stainless steel
and will deteriorate washer.