Operator Manual
Cycle and Control Value Programming
Table 5-1. Basil 9502 Cage and Rack Washer – Cycle Description Chart
ADJUSTABLE CYCLE PARAMETERS
PRE-WASH
WASH 1
M
ID-RINSE
WASH 2
R
INSE
FINAL RINSE
VAPOR EXHAUST
DRY
(OPTIONAL)
(OPTIONAL)
SELECTIONS
NON-RECIRCULATED
RECIRCULATED
NON-RECIRCULATED
RECIRCULATED
NON-RECIRCULATED
RECIRCULATED
NON-RECIRCULATED
SELECTABLE TANKS
RINSE
ALKALINE OR ACID
R
INSE
ACID OR ALKALINE
RINSE
RINSE
INJECTION RANGE
0 OZ/GAL- 4 OZ/GAL
0 OZ/GAL- 4 OZ/GAL
(0 ML/L - 32 ML/L)
(0 ML/L - 32 ML/L)
DURATION RANGE
00:20
00:00 - 10:00
00:20
00:00 - 10:00
00:20
00:00 - 08:00
01:00 - 05:00
00:00 - 15:00
LOAD TEMPERATURE
120°F - 165°F
120°F - 165°F
¤
160°F-185°F
140°F - 180°F
RANGE
(49°C - 74°C)
(49°C - 74C)
(71°C - 85°C)
(60°C - 82°C)
TANKS REFRESH RATE
15 % TO 100 %
15 % TO 100 %
PRE-PROGRAMMED CYCLES FACTORY SETTING
S
PRE-WASH
WASH 1
M
ID-RINSE
WASH 2
R
INSE
FINAL RINSE
VAPOR EXHAUST
DRY
SELECTED TANK
RINSE
ALKALINE
RINSE
IF SELECTED
RINSE
RINSE
INJECTION RATE
1 OZ/GAL (8 ML/L)
IF SELECTED
RODENTS
DURATION
00:20
02:00
IF SELECTED
IF SELECTED
IF SELECTED
03:00
01:00
15:00
NO
NO
NO
NO
TEMPERATURE
130°F (54°C)
IF SELECTED
185°F (85°C)
180°F (82°C)
TANK REFRESH RATE
15%
IF SELECTED
SELECTED TANK
RINSE
ALKALINE
RINSE
IF SELECTED
RINSE
RINSE
INJECTION RATE
1 OZ/GAL (8 ML/L)
IF SELECTED
BOTTLES
DURATION
00:20
02:00
IF SELECTED
IF SELECTED
IF SELECTED
03:00
01:00
15:00
NO
YES
NO
YES
TEMPERATURE
130°F (54°C)
IF SELECTED
185°F (85°C)
180°F (82°C)
TANK REFRESH RATE
15%
IF SELECTED
SELECTED TANK
RINSE
ALKALINE
RINSE
ACID
R
INSE
RINSE
INJECTION RATE
1 OZ/GAL (8 ML/L)
2 OZ/GAL (16 ML/L)
RABBITS
DURATION
00:20
02:00
IF SELECTED
08:00
IF SELECTED
02:00
01:00
10:00
YES
NO
YES
NO
TEMPERATURE
130°F (54°C)
160°F (71°C)
185°F (85°C)
180°F (82°C)
TANK REFRESH RATE
15%
15%
SELECTED TANK
RINSE
ALKALINE
RINSE
ACID
R
INSE
RINSE
INJECTION RATE
1 OZ/GAL (8 ML/L)
2 OZ/GAL (16 ML/L)
PRIMATES
DURATION
00:20
06:00
IF SELECTED
06:00
IF SELECTED
02:00
01:00
10:00
YES
NO
YES
NO
TEMPERATURE
130°F (54°C)
160°F (71°C)
185°F (85°C)
180°F (82°C)
TANK REFRESH RATE
15%
15%
SELECTED TANK
RINSE
IF SELECTED
RINSE
IF SELECTED
RINSE
RINSE
CYCLE 5
INJECTION RATE
IF SELECTED
IF SELECTED
TO
DURATION
00:20
IF SELECTED
IF SELECTED
IF SELECTED
IF SELECTED
IF SELECTED
01:00
IF SELECTED
NO
NO
NO
NO
CYCLE 10
TEMPERATURE
IF SELECTED
IF SELECTED
IF SELECTED
IF SELECTED
TANK REFRESH RATE
15%
IF SELECTED
DESCALER
DESCALER cycle can not be altered.
DRAIN
This cycle enables draining of TANKS 1, 2 & 3 (Optional), Purge Spray Headers
¤
Final Rinse temperature can be modified in Service Mode only. Modifying Final Rinse temperature can impact on the efficiency o
f the treatment. STERIS does not recommend performing a Thermal Rinse at a temperature below 185°F (85°C).
REVISION 0
REF.: #920-016-094
YES OR NO
YES OR NO
YES OR NO
YES OR NO
FLOOR TILTING
LOAD CHAMBER
FLOOR TILTING
MANIFOLD COUPLING
UNLOAD CHAMBER
UNLOAD CHAMBER
MANIFOLD COUPLING
LOAD CHAMBER
FLOOR TILTING
MANIFOLD COUPLING
FLOOR TILTING
MANIFOLD COUPLING
AUTOMATIC
LOAD CHAMBER
LOAD CHAMBER
UNLOAD CHAMBER
UNLOAD CHAMBER
DETAILS
5 MINUTES EACH TANK , 180°F MAINTAINED, NON GUARANTEED
1 MINUTE
DESCALER INJECTION IN EACH CHAMBER
5 MINUTES EACH TANK , 180°F MAINTAINED, NOT GUARANTEED
3 MINUTES EACH TANK , 120°F
1 MINUTE EACH TANK
MANUAL
CONTROL INDICATES OPERATOR TO ADD DESCALER IN EACH CHAMBER FOR EACH TANK
5 MINUTES EACH TANK , 180°F MAINTAINED, NOT GUARANTEED
3 MINUTES EACH TANK, 120°F
5 MINUTES EACH TANK , 180°F MAINTAINED, NON GUARANTEED
5 - DRAIN ALL TANKS
6 - FILL TANKS 1 & 2 ( 3 optional )
7 - RECIRCULATE EACH TANK
8 - VAPOR EXHAUST
3 - FILL THE SUMP WITH TANKS 1, 2 or (3 optional)
4 - RECIRCULATE EACH TANK
PHASES
1 - DRAIN TANKS 1 & 2 ( 3 optional ) & PIPING
2 - FILL TANKS 1 & 2 ( 3 optional )