6-7
Routine Maintenance
Operator Manual
920017-148
#920-502-066
1.
Remove end spray jets from each horizontal spray header, as indicated on
Figure 6-1.
2.
Use a fine wire (approximately the wire gauge of a paper clip) to clear
sediment from all spray nozzles. Rinse under running water to clean out
sediment.
3.
Process a DRAIN Cycle including a Purge Travelers phase (see Section 4.6,
DRAIN Cycle) with spray jets removed.
4.
Reinstall end spray jets on each horizontal spray headers. Assure jets are
reinstalled in proper horizontal position.
5.
Remove end caps from horizontal spray headers to visually inspect for
debris and sediments. Clean if necessary.
6.
Reinstall end caps.
1.
Remove Rotary Spray Holder and Rotary Spray Arms from Accessory.
2.
Unscrew spray arms from Spray Arm Hub. Rinse Spray Arms under running
water to clean out sediment.
3.
Remove End Spray Jets from Spray Arms. Use a fine wire (approximately
the wire gauge of a paper clip) to clean out sediment. Rinse under running
water.
4.
Reinstall spray jets on Rotary Spray Arms.
5.
Place bushing on top of Rotary Spray Arm Hub and reinstall Rotary Spray
Arm assembly to Centra Header Manifold.
6.2.4 Central Header
Manifold (Option)
6.2.3 Horizontal Spray
Headers and Spray Jets
Figure 6-1. Spray Jets and End Caps
W A R N I N G – S L I P P I N G
HAZARD: To avoid slippery
floor conditions, immedi-
ately wipe up any spilled
liquids or condensation.
W A R N I N G – S L I P P I N G
HAZARD: To avoid slippery
floor conditions, immedi-
ately wipe up any spilled
liquids or condensation.
Remove This
End Spray Jet
End
Cap
End
Cap
Remove This
End Spray Jet