6-13
Routine Maintenance
Operator Manual
920017-148
6.
When all tanks are completely drained, tanks start filling with clean water.
Display shows:
DESCALER
ALKALINE
FILLING
MM:SS
... then:
DESCALER
RINSE
FILLING
... then:
DESCALER
ACID
FILLING
7.
When all tanks are filled with clean water, cycle stops and display shows:
ADD DESCALER IN BOTH
CHAMBERS-PRESS START
8.
On load side, press
DOOR OPEN
touch pad to open load door.
9.
Pour 76 oz (2.25 L) of Liquid Descaler Liquid Scale Remover into load
chamber sump.
10. Press
DOOR CLOSE
touch pad to close load door.
11. On unload side press
DOOR OPEN
touch pad to open unload door.
12. Pour 76 oz (2.25 L) of Liquid Descaler Liquid Scale Remover into unload
chamber sump.
13. Press
DOOR CLOSE
touch pad to close unload door.
14. Press
CYCLE START
touch pad to run DESCALER cycle. Water from
sumps will be sent to Alkaline Tanks to initiate descaling. Display shows:
DESCALER
ALKALINE
PRIMING
WARNING – CHEMICAL
BURN AND/OR EYE INJURY
HAZARD: Wear appropriate
personal protective equip-
ment (PPE) when using a
descaling product. Avoid
contact with eyes or skin. If
spilled or splashed, flush
with water for 15 minutes. If
swallowed, DO NOT induce
vomiting. Administer an al-
kali with plenty of water.
Seek medical attention im-
mediately.
DOOR
OPEN
DOOR
CLOSE
CYCLE
START
DOOR
OPEN
DOOR
CLOSE