6-18
920017-148
Operator Manual
Routine Maintenance
6.4 Chemical
Container
Replacement
To achieve maximum cleaning efficiency, select a chemical appropriate to soil
type being processed. See Section 2.3,
Chemical Additive Specifications, and
always follow chemical manufacturer's recommendations for amount of chemi-
cal to be used.
NOTE: Always use a non-foaming chemical for effective cleaning and proper
pump operation.
When chemical is low or has run out:
1. Remove pickup tube and level sensor assembly from empty container and
remove container.
2. Install a new container.
3. Insert pickup tube and level sensor vertically into new container.
NOTE: System is self-priming, therefore no other operation is necessary.
WARNING – CHEMICAL
BURN AND/OR EYE INJURY
HAZARD:
•
Always use non-foaming
chemical for effective clean-
ing and proper pump and
water level control opera-
tion. Follow manufacturer's
recommendations for
amount of chemical to be
used.
•
Detergents are caustic and
can cause adverse effects to
exposed tissues. Do not get
in eyes, on skin or attempt
to swallow. Read and follow
the precautions and instruc-
tions on the detergent label
and in the Material Safety
Data Sheet (MSDS) prior to
handling detergent contain-
ers, or servicing the deter-
gent injection pumps and
lines. Wear appropriate per-
sonal protective equipment
(PPE) whenever handling
chemicals or servicing
chemical injection pumps
and lines.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
When choosing a chemical,
select one with a low chlo-
ride content. Chloride is
harmful to stainless steel
and will deteriorate washer.
Pickup Tube
Level Sensor
Chemical
Container
Figure 6-7. Chemical Container