5-10
920017-148
Operator Manual
Cycle and Control Value Programming
46. The load MANIFOLD COUPLING screen appears. Display shows:
LOAD MANIFOLD ?
NO
__indicates flashing position.
NOTE: The load MANIFOLD COUPLING display is not shown if load
AUTOMATIC FLOOR TILTING option is activated.
IMPORTANT:
This screen is to activate or deactivate manifold coupling
for the load chamber only.
Press
VALUE
(up or down) touch pad to change YES to NO. Press
CHANGE VALUES
touch pad to confirm selection.
47.
If unit is equipped with Automatic Floor Tilting option
, the unload
AUTOMATIC FLOOR TILTING screen appears. Display shows:
UNLOAD FLOOR TILT ?
NO
__indicates flashing position.
IMPORTANT:
This screen is to activate or deactivate automatic floor
tilting for the unload chamber only.
Press
VALUE
(up or down) touch pad to change YES to NO. Press
CHANGE VALUES
touch pad to confirm selection.
48. The unload
MANIFOLD COUPLING
screen appears. Display shows:
UNLOAD MANIFOLD ?
NO
__indicates flashing position.
NOTE: The unload MANIFOLD COUPLING display is not shown if unload
AUTOMATIC FLOOR TILTING option is activated.
IMPORTANT:
This screen is to activate or deactivate manifold coupling
for the unload chamber only.
Press
VALUE
(up or down) touch pad to change YES to NO. Press
CHANGE VALUES
touch pad to confirm selection.
Display shows, for
example:
DATE
YY/MM/DD
__indicates flashing position.
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼
CHANGE
VALUES
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: Do
not process load using Cen-
tral Header Manifold or
Bottle Washing Cart in a
chamber where Automatic
Floor Tilting option is acti-
vated. If Automatic Floor
Tilting is used, manifolded
water inlet and washer will
be damaged.
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼
CHANGE
VALUES
CHANGE
VALUES
VALUE
▲
▲
▲
▲
▲ ▼
▼
▼
▼
▼