5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.4 Vacuum system
Operating Instructions Simulation Unit SIM MODULAR
5. Press the button.
The mixture of water and cleaning agent (optional) is pumped
into the suction hose cleaning container and drawn off by the
suction hoses. An acoustic signal will sound after completing the
suction hose cleaning.
6. Then place the suction hoses back into their holders on the assistant
element.
7. Repeat this procedure for the other suction handpieces.
Vacuum system, hose fittings
Refilling the cleaning agent
When the adjacent display appears on the EasyPad display or in the
status column of the touchscreen, the cleaning agent for cleaning the
suction system is almost finished. It should then be refilled as soon as
possible.
Vacuum system, Mixing with DENTOSEPT P
1. Open the maintenance flap on the base of the supply module. The
cleaning agent tank for chemical suction hose cleaning is located on
the left side.
2. Pull the tank slightly out of the supply module. At the same time,
check the hose line.
3. Open the lock of the tank and add the cleaning agent. The cleaning
agent tank has a capacity of 1 liter.
4. Close and slide the tank back into the supply module after filling. The
filling opening must face the front.
If the simulation unit is equipped with this option, the drains should be
cleaned weekly.
CAUTION
Suction hose cleaning must not be interrupted to ensure that no residual
water remains in the container.
CAUTION
Regularly disinfect the hose fittings on the supply module.
!
Aqua
Amalg
San
1d
41d
B
10:45:03
chem
CAUTION
The cleaning agent for the vacuum system could be mistaken for the
disinfectant for the water lines.
➢ Do not fill DENTOSEPT P disinfectant into the cleaning agent tank
for chemical suction hose cleaning! Use only "Agents for the suction
lines" approved by Sirona, see Care, cleaning and disinfecting
NOTICE
Common household cleaning agents produce foam.
This results in foam and water being sucked into the dry suctioning
system, which can cause damage to the suction machine.
➢ Only use Sirona approved care, cleaning and disinfecting agents
(see Care, cleaning, and disinfecting agents [ → 148]).
1LCleanser