5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.10 Sanitizing
båÖäáëÜ
Starting sanitation
➢ Touch the
Start
key.
The sanitation process starts.
Error message: Refill with DENTOSEPT P disinfectant
If the
Disinf
display appears after the sanitation process is started, the
supply of DENTOSEPT P disinfectant in the reservoir of the supply
module is not sufficient to sanitize the simulation unit. Sanitation cannot
be started with insufficient disinfectant; see "Refilling with
DENTOSEPT P disinfectant".
Error message: Deposit instruments
If the
Deposit instruments
display appears after sanitization has been
started, the simulation unit has detected that not all of the instruments
have been placed in their holders.
➢ Check the instruments marked with a warning triangle on the
touchscreen.
When all of the instruments have been deposited, sanitation
phase 1 begins automatically.
Sanitation phase 1 – Inserting instruments into the cleaning adapters
on the supply module
The supply module has integrated adapters for water-carrying
instruments. This allows all instruments to be simultaneously treated with
a high concentration of sanitation solution and then rinsed with water. For
this purpose, sanitizable instruments must be inserted into the adapters
on the supply module.
✔
Sanitization phase 1 is highlighted on the touchscreen.
✔
The
Insert instruments in water unit
display appears on the
touchscreen.
1. Remove the SPRAYVIT sleeves from the valve bodies and the
straight and contra-angle handpieces from the sanitizable
instruments.
Start
Start
1
2
3
4
5
6
H2O
Desinf
Stop
Desinf
Start
1
2
3
4
5
6
H2O
Desinf
Stop
!
!
Start
1
2
3
4
5
6
H2O
Desinf
Start