4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.5 Dentist element
båÖäáëÜ
4.5.2
Positioning the dentist element
Dentist element on support arm, collision, manual movement
Dentist element, sudden movements
The dentist element is attached to the simulation unit with a flexible
support arm. The dentist element is held in place at the adjusted height
with a pneumatic locking brake in the support arm.
The hand knob to adjust the support arm brake is underneath the flexible
support arm.
The knob must only be adjusted if required.
➢ Hold the hand knob in your hand and turn it counterclockwise to
release the brake, or clockwise to tighten the brake.
There is a push button near the two handles on the sides of the CS dentist
element to release the brake. The TS dentist element only has one button
on the right side.
The button must only be pressed to move the chair vertically. Horizontal
movements are also possible without releasing the brake.
1. Grip a handle with your hand and keep the button pressed.
The locking brake is released with the sound of pressurized air.
The dentist element can be raised and lowered.
2. Position the dentist element and release the button.
The dentist element is secured at the height set.
Depending on the equipment of the dentist element, the retention force of
the locking brake can be changed by the technician.
WARNING
It is possible to position the dentist element in the operating area of the
user.
Moving the patient simulator can result in collision with the dentist
element or its support arm. There is the danger of injury due to crushing.
➢ Before moving the patient simulator, position the dentist element to
make a collision impossible.
NOTICE
Sudden movements can cause instruments to fall out of the holder in the
dentist element.
➢ Try to avoid sudden movements of the dentist element.