5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.3 Instruments and instrument hoses
Operating Instructions Simulation Unit SIM MODULAR
Purging water lines
✔
The
Purge function activated
display is shown on the touchscreen.
1. Remove the instruments from the TS dentist element that are to be
purged and hold them over a bucket. For
the CS dentist element, move the swivel arms of the instruments to
be purged into their operational position so that the weight of the
instrument hoses keeps them in this position.
After an instrument is removed, the purge process starts
automatically after about 5 seconds (except for the 3-way
syringe). The removed instruments are purged with water for the
duration of the set purging time.
2. Place the removed instruments back in their holders after the purging
process.
3. Hold the 3-way syringe of the dentist and assistant element over the
bucket and press the water key for at least 20 seconds.
4. Touch the
Stop
key to end the purge program.
After all instruments have been put back in their holders, the
Purge function activated
display disappears. The
Start dialog
is
shown.
The water line purging procedure is finished. The simulation unit is
ready for operation again.
Stop
CAUTION
Do not purge more than two burs simultaneously.
If the simulation unit features the option of a third burr drive, the water
flow rate is reduced when purging simultaneously.
➢ Do not remove more than two burr drives for purging. Purge the third
burr drive separately afterwards.
Stop