157
Sirona Dental Systems GmbH
5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.3 Instruments and instrument hoses
båÖäáëÜ
Error message: Deposit instruments
If the
Deposit instruments
display appears after the purge function has
been started, the simulation unit has detected that not all of the
instruments have been placed in their holders.
➢ Check the instruments marked with a warning triangle on the
touchscreen.
When all of the instruments have been put in the holders, the
purge program will continue.
Error message: Refill water (only for stand-alone water supply)
If the
Refill water
display appears after the purge function has been
started, there is not sufficient water in the disinfectant tank of the supply
module to purge the water lines. The purge function cannot be started
with insufficient water.
➢ Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the supply module.
When sufficient water has been refilled, the purge program
continues.
Setting the purge time and starting the purge function
The purge time of the removed instruments can be set between 20 and
120 seconds.
✔
The
Purge
dialog is displayed on the touchscreen.
1. Use the – and + keys to set the purging time.
2. Touch the
Start
key.
120 s
Start
Stop
!
!
120 s
Start
Stop
Aqua
120 s
Start
Start