4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.5 Dentist element
båÖäáëÜ
4.5.11.3
Setting the intensity on the EasyTouch
Intensities are saved on the three favorite keys that can be accessed by
pressing them. Individual intensities can be saved on the favorite keys for
each user profile.
On the C+ electronic foot switch, spray activation is saved on the favorite
keys. The settings can also be made via the 4-way switch plate.
Accessing intensity settings
✔
The SIROSONIC L scaler is removed.
✔
The
Ultrasonic dialog
is displayed on the touchscreen.
➢ Touch one of the favorite keys in the bottom line briefly (< 1 s).
The favorite key is highlighted orange. The selected intensity is
displayed to the left of the favorite keys.
By moving the 4-way switch plate of the C+ electronic foot switch to the
right, the settings of the favorite keys can be accessed consecutively.
Changing the intensity
➢ Press and hold the left or right favorite key (> 1 s).
The intensity is increased or reduced. The favorite keys are
shaded gray for intermediate values.
The intensity can also be increased or reduced by moving the 4-way
switch plate of the C+ electronic foot switch up or down.
Saving intensity
✔
The desired intensity is set.
✔
Via the C+ electronic foot switch: The spray is switched on or off (will
be saved).
1. Press and hold the middle favorite
Set
(> 2 s).
The intensity flashes on the touchscreen.
2. Press one of the three favorite keys.
An acoustic signal sounds. The intensity set and, if applicable,
activation of spray will be saved on the favorite key. The selected
intensity is displayed on the favorite key.
While the intensity flashes on the touchscreen, pressing any other key on
the touchscreen will interrupt saving to a favorite key.