5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.10 Sanitizing
Operating Instructions Simulation Unit SIM MODULAR
3. If the simulation unit is equipped with the option to switch to the public
drinking water supply, it must be switched to operation with the stand-
alone water supply from the fresh water bottle for sanitizing. Turn the
lever clockwise to the bottle symbol. For more information, see the
section “Water supply with fresh water bottle [ → 107].”
4. Unscrew the fresh water bottle from the holder on the support tube
and empty it completely.
5. Fill the fresh water bottle with approx. 0.4 to 0.5 liters of pure
DENTOSEPT P.
6. Attach the hose to the fresh water bottle and screw the bottle into the
holder.
Displaying the sanitizing operating context
✔
All instruments are in their holders.
1. Press the
Torque control/Purge
fixed key.
The
Purge
operating context is shown on the EasyPad display.
2. Press favorite key 3.
The
Sanitizing
operating context is displayed.
3. Press the favorite key
2 / Setup
.
A revolving element is displayed on the EasyPad next to the text
“dES.”
Treating water lines with DENTOSEPT P disinfectant
The water lines of the instrument hoses and SPRAYVIT hoses are filled
with pure DENTOSEPT P.
1. Hold the 3-way syringe of the dentist and assistant element over a
bucket and press the water key for at least 15 seconds.
Purge
3
2
Setup