5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.10 Sanitizing
båÖäáëÜ
All instruments which can be sanitized are inserted into the
cleaning adapters.
6. Touch the
Start
key on the touchscreen.
Sanitation phase 2 – Checking the water flow
First, the simulation unit checks whether there is water flow through the
instruments.
✔
Sanitization phase 2 is highlighted on the touchscreen.
➢ Wait briefly until the water flow has been checked.
If sufficient water flow is present, the simulation unit continues
with sanitation phase 3.
Error message: No water flow
If the simulation unit detects no water flow through an instrument, you can
try to restore water flow. If this is not possible, the instrument concerned
can be excluded from the sanitation process.
1. Check the water flow through the instruments in the instrument
positions marked with a warning triangle on the touchscreen. Set the
instruments to maximum water flow. Leave all instruments plugged
into the supply module.
If the simulation unit detects the water flow, the warning triangle
will disappear. If there is sufficient water flow through all
instruments, sanitation phase 3 automatically continues.
2. If you want to exclude the instruments concerned from sanitation,
touch the
Exclude instrument
key.
The simulation unit continues with sanitation phase 3. The water
lines you excluded are not included in the sanitation process.
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
If the hoses are pinched, the water flow will be obstructed during
purging.