5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.4 Vacuum system
båÖäáëÜ
Rinsing the cleaning agent
1. Following the cleaning process, pour at least 1 liter of water into the
container.
The water is suctioned off, thus preventing any cleaning agent
residues from remaining in the suction hoses.
2. After finishing the aspiration process, place the suction hoses back in
their holders.
If the simulation unit is equipped with a third suction hose, repeat the
procedure.
If the simulation unit is equipped with a wet suction device, the filter insert
of the wet suction device must also be cleaned once a month after
cleaning the vacuum system (see “Clean the filter insert of the wet suction
device [ → 183]”).
If the simulation unit is equipped with an air jet pump (Air Venturi), the
collection container should be emptied on a weekly basis after cleaning
the suction system (see “Emptying the collector of the air jet pump
1l H
O
2
max.