4 Operation
Sirona Dental Systems GmbH
4.7 Stand-alone water supply
Operating Instructions Simulation Unit SIM MODULAR
4.7.2
Water supply through the disinfection system
In the simulation unit, the stand-alone water supply should be set only as
a temporary operating state if the public drinking water is contaminated
and not as a permanent operating state.
Manually sanitize water supply, stand-alone treatment center
Stand-alone water supply, filling with distilled water
4.7.2.1
Switching to stand-alone water supply on the EasyPad
Accessing the operating context of the stand-alone water supply
✔
All instruments are in their holders.
1. Press the
Torque control/Purge
fixed key.
The
Purge
operating context is shown on the EasyPad display.
2. Press favorite key 3 three times.
The
Stand-alone water supply
operating context is displayed.
If there is a "0" next to the text "AqU," the simulation unit takes the water
from the public drinking water supply. If there is a "1," the simulation unit
is switched to the stand-alone water supply. To change the mode of
operation, proceed as follows:
Changing disinfection system to stand-alone water supply
1. Press the favorite key
2 / Setup
.
The selection focus flashes. The simulation unit is set for
operation with the public drinking water supply.
2. Press favorite key 3.
3. A "1" flashes in the selection focus. The simulation unit should be
switched to stand-alone water supply.
CAUTION
If the stand-alone water supply is used more than 28 days in exceptional
cases, the simulation unit must be sanitized.
For details on the sanitizing procedure, please see the section
NOTICE
Always fill distilled food-grade water (< 100 CFU/ml), never drinking
water e.g., from beverage bottles (due to the minerals contained
therein), into the disinfectant tank of the simulation unit.
Aqua purificata or aqua destillata can be used as distilled water.
Purge
3
2
Setup
3