4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.9 SIVISION digital video system
båÖäáëÜ
4.9.1
SIVISION monitor
22" AC monitor
The monitor is equipped with loudspeakers. There is an IEC socket on the
back of the monitor for the power supply. Please refer to the operating
instructions for the 22" AC monitor for details.
The monitor is mounted on the lamp support rod. It can be rotated and
swiveled.
Monitors that are equipped with four threaded holes at a distance of 75
mm on the rear can be mounted on the monitor holder on the LEDview
lamp support rod (VESA standard). Monitors with threaded holes at a
distance of 100 mm can also be mounted on the holder of the LEDview
lamp support rod.
Monitor, not suitable without approval
Connection of loudspeakers
WARNING
Monitors without approval must not be connected.
They endanger the product safety of the simulation unit.
➢ Use only monitors that are approved according to IEC 60950-1
(office equipment) or IEC 60601-1 (medical devices).
WARNING
Unsuitable devices can be connected to the loudspeaker port of the
monitor.
The connection of unsuitable devices endangers the product safety of
the simulation unit.
➢ The loudspeaker port of the monitor may be connected only to a
device that complies with IEC 60950-1 (office equipment such as
PCs) or IEC 60601-1 (medical devices). Under no circumstances
should it be connected e.g. to a stereo system, etc.