4 Operation
Sirona Dental Systems GmbH
4.7 Stand-alone water supply
Operating Instructions Simulation Unit SIM MODULAR
4.7.2.2
Switching to stand-alone water supply with the EasyTouch
Opening the Start sub-screen
✔
All instruments are in their holders.
✔
The
Start dialog
is displayed on the touchscreen.
➢ Press the
Sub-dialog
fixed key.
The
Start
sub-dialog is displayed.
Changing disinfection system to stand-alone water supply
If the
Aqua
key is displayed gray in the
Start
sub-dialog, the simulation
unit gets the water from the public water supply (municipal water). If the
key is highlighted orange, the simulation unit is switched to the stand-
alone water supply. To change the mode of operation, proceed as
follows:
1. Press and hold the
Aqua
key (> 2 s).
The
Aqua
key flashes orange until the disinfectant tank is
emptied. Moreover, the
Changing to self-sufficient water supply
display appears.
If there is still disinfectant in the tank, it is rinsed down the drain.
This can take up to 6 minutes. The
Refill water
display then
appears on the touchscreen.
2. Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the simulation unit. The tank has a capacity of
approx. 1.3 liters. It is full when the water surface is visible on the filter
of the filling funnel.