2 Safety instructions
Operating Instructions Simulation Unit SIM MODULAR
2.3 On-site installation
båÖäáëÜ
2.3
On-site installation
The on-site installation must have been performed according to our
requirements. The details are described in the document "Installation
Requirements".
2.4
Installing the simulation unit
Installation must be carried out by authorized personnel according to the
installation instructions.
2.5
Media quality
The air and water supplies must meet the requirements specified in the
installation requirements. In addition, the optional fresh water bottle or
empty disinfectant tank can be used as a water container.
Use only drinking water and dry oil-free hygienically clean air in this
simulation workstation. As the owner of the simulation workstation, you
are generally responsible for the water quality.
Please note that drinking water, e.g., water from bottles, must never be
filled into the disinfectant tank of the simulation workstation (due to the
minerals contained therein). Aqua purificata or aqua destillata can be
used as distilled water. Water from bottles can be used without any
restrictions for the fresh water bottle.
CAUTION
Microorganisms can multiply in the water.
These microorganisms could increase the risk of damage to one’s
health.
➢ Sirona recommends to never operate the simulation unit without
disinfectant.
➢ Mix fresh water for the stand-alone water supply every day. At the
end of the work day, the fresh water bottle must be emptied.