4 Operation
Sirona Dental Systems GmbH
4.5 Dentist element
Operating Instructions Simulation Unit SIM MODULAR
4.5.11
SIROSONIC L scaler
The SIROSONIC L scaler is used for removing plaque and rinsing the root
canal for endodontic treatments.
Ultrasonic handpiece, observe the operating Instructions
4.5.11.1
Safety instructions
The torque wrench is used as a tool for screwing in instrument tips and,
at the same time, to protect against injury.
Ultrasonic handpiece, torque wrench
Ultrasonic handpiece, only Sirona tips
4.5.11.2
Setting the intensity on the EasyPad
Intensities are saved on the favorite keys 1, 2, and 3, which can be
accessed by pressing them. Individual intensities can be saved on the
favorite keys for each user profile.
On the C+ electronic foot switch, spray activation is saved on the favorite
keys. The settings can also be made via the 4-way switch plate.
Accessing intensity settings
✔
The SIROSONIC L scaler is removed.
✔
The intensity setting is shown on the EasyPad display.
➢ Press the favorite keys briefly (< 1 s).
The intensity saved on the favorite key is shown as a percentage
on the EasyPad display.
By moving the 4-way switch plate of the C+ electronic foot switch to the
right, the settings of the favorite keys can be accessed consecutively.
IMPORTANT
See also the SIROSONIC L operating instructions.
CAUTION
Ultrasonic tips are sharply pointed.
There is a risk of injuring one's hand on the deposited scaler.
➢ Always attach the torque wrench to the scaler for protection as soon
as you deposit the handpiece.
CAUTION
In some cases, ultrasonic tips from other manufacturers do not
guarantee safe operation.
➢ Use only ultrasonic tips from Sirona.