5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.3 Instruments and instrument hoses
båÖäáëÜ
5.3.2.1
Purge function on the EasyPad
Setting the purging time
The purging time of the removed instruments can be set between 20 and
120 seconds in the setup for the simulation unit, see “Setting the purging
time for the purge function” [ → 138].
Switching on the purging function
✔
All instruments are in their holders.
1. Press the
Torque control/purge
key.
The
Purge
operating context is shown on the EasyPad display.
2. Press the favorite key
2 / Setup
.
Error message: Refill water (only for stand-alone water supply)
If the
Refill water
display appears on the EasyPad display after the purge
function has been started, there is not sufficient water in the disinfectant
tank of the supply module to purge the water lines. The purge function
cannot be started with insufficient water.
➢ Mix distilled water with the DENTOSEPT P disinfectant at a ratio of
100:1 (1 liter of water, 10 ml DENTOSEPT P) and fill this into the
disinfectant tank of the supply module.
When sufficient water has been refilled, the purge program
continues.
Purging water lines
✔
A revolving element is shown on the EasyPad display next to the
Purge
display.
1. Remove the instruments from the TS dentist element that are to be
purged and hold them over a bucket. For
the CS dentist element, move the swivel arms of the instruments to
be purged into their operational position so that the weight of the
instrument hoses keeps them in this position.
After an instrument is removed, the purge process starts
automatically after about 5 seconds (except for the 3-way
syringe). The removed instruments are purged with water for the
duration of the set purging time.
Purge
2
Setup
CAUTION
Do not purge more than two burs simultaneously.
If the simulation unit features the option of a third burr drive, the water
flow rate is reduced when purging simultaneously.
➢ Do not remove more than two burr drives for purging. Purge the third
burr drive separately afterwards.