5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.10 Sanitizing
Operating Instructions Simulation Unit SIM MODULAR
All water-carrying instruments are inserted into the cleaning
adapters.
6. Press the favorite key
2 / Setup
.
The sanitizing program will start. A revolving element is shown on
the EasyPad display next to the text "dES".
Treating water lines with DENTOSEPT P disinfectant
The water lines of the instrument hoses and SPRAYVIT hoses are filled
with pure DENTOSEPT P.
Tumbler filling, discolorations due to DENTOSEPT P
Allow 24 hours reaction time
In order to effectively combat germs, you must allow the DENTOSEPT P
disinfectant to act at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
➢ Switch the simulation unit off using the main switch on the supply
module. Make sure that the simulation unit is switched off for at least
24 hours, but not longer than 3 days (maximum sanitation time).
➢ Switch the simulation unit on again after 24 hours and before 3 days
have elapsed.
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
If the hoses are pinched, the water flow will be obstructed during
purging.
The water flow rate to the instruments is not checked by the simulation
unit.
2
Setup
NOTICE
Splashes of pure DENTOSEPT P can result in discoloration of surfaces
if left for too long.
Any splashes should therefore be removed immediately using a moist
cloth.
IMPORTANT
Blocking water and air supply
When the simulation unit is turned on again after 24 hours, the sanitizing
process continues automatically. However, if the water and air supplies
are switched off, the DENTOSEPT P disinfectant cannot be rinsed out
of the water lines.
Switch the simulation unit back on after 24 hours only when the water
and air supply lines are open.