5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.1 Basics
Operating Instructions Simulation Unit SIM MODULAR
5.1.2
Care, cleaning, and disinfecting agents
Approved care, cleaning, and disinfecting agents
Care, cleaning, and disinfecting agents, with REF number
For a continuously updated list of approved agents, please visit
"www.sirona.com"
. To access the online portal for technical
documentation, follow the
"SERVICE"
/
"Technical Documentation"
menu items in the navigation bar. The portal can also be accessed
directly via the following address http://www.sirona.com/manuals. Click
on the menu item
"General documents"
and then open the
"Care,
cleaning and disinfection agents"
document.
If you do not have any access to the Internet, please contact your dental
depot to order the list (REF 59 70 905).
5.1.3
Conduct a microbiological water test
Perform microbiological tests of the water from the simulation unit at
regular intervals and after longer periods of disuse > 1 week (see also
“Media quality”). Start the tests in intervals of no more than two weeks and
adjust the time intervals depending on the results. In addition to running
laboratory tests, you can also use the “Total Count Tester” as a simple
means of performing this test.
To order the motor total count tester, see "Spare parts and consumables".
Total Count Tester, shelf life
The cardboard disk contains a dehydrated culture medium. It is activated
by the sample and serves as culture medium for a number of bacteria.
The bacterial count provides information on the hygiene quality of the
water.
Be careful not to touch the inside of the test container or the part of the
tester to be immersed (with nutrient medium) prior to incubation.
1. Press the tumbler filler for at least one minute.
2. Allow water to run out of the 3-way syringe in the bucket for about 1
minute.
3. Use the 3-way syringe to fill cold water into the test container up to its
upper mark.
4. To neutralize the disinfecting agent of the water sample, add approx.
1.5 g of fixing salt (sodium thiosulfate). Fixing salt can be obtained in
pharmacies or from chemical dealers.
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning and disinfecting agents approved by Sirona!
IMPORTANT
Shelf life of the total count tester
The maximum shelf life of the total count tester is 1 year after the date
of receipt.