7ML19985XC81
SITRANS LR200 (HART) – QUICK START MANUAL
Page EN-19
m
mmmm
Englis
h
c)
Press RIGHT arrow
to open Edit mode or DOWN arrow
to accept default
values and move directly to the next item.
d)
To change a setting, scroll to the desired item or key in a new value.
e)
After modifying a value, press RIGHT arrow
to accept it and press DOWN arrow
to move to the next item.
f)
Quick Start settings take effect only after you select Yes to Apply changes in
step 1.7.
1.1. Material
1.2. Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
1.3. Sensor Units
Selects the units for the Quick Start variables (high and
low calibration point, and level, distance, or space).
1.4. Operation
Options
LIQUID
LIQUID LOW DK (low dielectric liquid)
Options
SLOW
0.1 m/minute
MED
1.0 m/minute
FAST
10.0 m/minute
Operation
types
NO
SERVICE
The SITRANS LR200 stops updating measurements and associ-
ated loop current. Last valid measurement is displayed.
LEVEL
Distance to material surface referenced from Low Calibration
Point (process empty level).
SPACE
Distance to material surface referenced from High Calibration
Point (process full level).
DISTANCE
Distance to material surface referenced from Sensor Reference
Point.
Options
M, CM, MM, FT, IN
high cal.
point
low cal.
point
level
space
distance
sensor
reference
point
Summary of Contents for Sitrans LR200
Page 1: ...Quick Start Manual May 2008 LR200 HART sitrans ...
Page 25: ...Page EN 24 SITRANS LR200 HART QUICK START MANUAL 7ML19985XC81 mmmmm English Notes ...
Page 49: ...Side DK 24 SITRANS LR200 HART QUICK START MANUAL 7ML19985XC81 mmmmm Dansk Bemærkninger ...
Page 73: ...Seite DE 24 SITRANS LR200 HART KURZANLEITUNG 7ML19985XC81 mmmmm Deutsch Notizen ...
Page 74: ...7ML19985XC81 SITRANS LR200 HART KURZANLEITUNG Seite DE 25 mmmmm Deutsch ...
Page 75: ...Seite DE 26 SITRANS LR200 HART KURZANLEITUNG 7ML19985XC81 mmmmm Deutsch ...
Page 123: ...Página ES 24 SITRANS LR200 HART ARRANQUE RÁPIDO 7ML19985XC81 mmmmm Español Notas ...
Page 147: ...Page FR 24 SITRANS LR200 HART MISE EN SERVICE RAPIDE 7ML19985XC81 mmmmm Français Remarques ...
Page 171: ...Pagina IT 24 SITRANS LR200 HART AVVIO RAPIDO 7ML19985XC81 mmmmm Italiano Nota ...
Page 196: ...7ML19985XC81 SITRANS LR200 HART QUICK START HANDLEIDING pagina NL 25 mmmmm Nederlands ...
Page 197: ...Pagina NL 26 SITRANS LR200 HART QUICK START HANDLEIDING 7ML19985XC81 mmmmm Nederlands ...
Page 245: ...Sivu SU 24 SITRANS LR200 HART PIKAKÄYTTÖOPAS 7ML19985XC81 mmmmm Suomi Huomautukset ...
Page 269: ...Sida SE 24 SITRANS LR200 HART SNABBSTARTSMANUAL 7ML19985XC81 mmmmm Svenska Anmärkningar ...