3-33
1020NFM-1D
Section 3
9. Realign left edge of the guide with the transducer edge mark. Line up both vertical edges of the
guide and ensuring that it is snug around the pipe, mark along the overlapping edge.
10. Remove Mylar Mounting Guide and lay it out on a flat surface. Either measure the exact distance
halfway between the overlap edge and the mark at the overlap, or fold the guide from the overlap
edge to overlap mark and draw a line at the fold or halfway point (see below).
Transducer
Pipe
Spacer Bar
Mylar
Spacing
Guide
Transducer
Spacer Bar
Dot
Dot
Line
A
B
C
Transducer
Spacer Bar
Spacer Bar
Dot
Line
Transducer
Mylar
Spacing
Guide
Transducer
Edge Line
Transducer
Dot
Pipe
WRAPPING THE MYLAR SPACING GUIDE AROUND THE PIPE
END VIEW
3" Overlapping
Edge
Mylar
Spacing Guide
Trim material from
inner edge if
necessary
Trim material from
inner edge if
necessary
3” Overlapping
Edge
Mylar
Spacing Guide
End View
FINDING THE HALFWAY DISTANCE
Overlap Edge
Mark (or fold) exactly
at half-way point
SPACING
GUIDE
CIRCUMFERENCE
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......