4-6
1020NFM-1D
Section 4
6. To move the highlight to [Install Flow Tube] press the <Down Arrow>.
7. To access the [Install Flow Tube] option list press the <Right Arrow>. Note that this list shows the
currently available flow tube sizes. The number is the size (e.g., 1). The letter represents the mate-
rial (e.g., S = Steel, P = PVC, T = Teflon, K = Kynar). In this example, we will install a size 1S Flow
Tube. This 1011FTP-1S (Stainless Steel) unit uses a 5/8” Swagelok Tube Fitting.
8. To select size
1S
press <Enter>. This moves the highlight to [Install Complete?].
Siemens 2 Channel [1]
FLOW1
Install FlowTube
1S
Install Complete?
No
Empty Set <FT>
Chan Not Setup
Zero Flow Adjust
Chan Not Setup
Pick Flow Tube size and Type
Install FlowTube
Siemens 2 Channel [1]
FLOW1
Install FlowTube >1S
Install Complete?
1P
Empty Set <FT>
1K
Zero Flow Adjust
1T
2S
2P
2K
2T
Pick Flow Tube size and Type
Install FlowTube
Note that the next menu down is Liquid Data. This menu allows you to enter Viscosity (in centistokes)
and the Density (specific gravity) data for the liquid to be measured. Data entry in this menu is
optional. The default values are those of Water at 68
°
F (20
°
C). If you choose to edit these param-
eters, you should be aware that the viscosity of the liquid is an important factor. It governs the degree
of Reynolds Number compensation that the flow computer applies to the final rate output. Therefore,
flow data errors could result if you enter inaccurate data. Note that the Siemens Technical Service
Group may be able to provide the viscosity data for the liquid to be measured.
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......