6-16
1020NFM-1D
Section 6
6.4
THE PICK/INSTALL XDCR MENU
Use this menu after creating a new site setup in the Channel Setup menu, and defining the pipe
parameters in the Pipe Data menu.
Based on pipe data (and optionally application data) entries, the Pick/Install Xdcr menu automatically
identifies the most suitable transducers for the application. It recommends the appropriate mounting
mode (direct or reflect) and lists the Spacer Bar or Mounting Track part number and spacing index.
Ideally, you will be able to use the primary recommendations. However, you can edit the menu entries
as required to accommodate different transducers or mounting configurations.
The flow computer will adjust its parameters to optimize performance based on your selections. The
[Ltn Value (in)] menu cell shows the required spacing distance (in inches or millimeters) between the
upstream and downstream transducers. Use the [Install Complete?] menu cell to inform the flow com-
puter that you completed the physical mounting of the transducers. You can define the empty pipe and
zero flow values once the transducers are operational.
Scroll List and select desired Model
Pick/Install Xdcr
Transducer Model
1011 Universal
Transducer Size
B3
Xdcr Mount Mode
Reflect
Spacing Offset
Spacer Bar PB-1
Number Index
0
Spacing Method
Track 1012TAH
Ltn Value
0.565
Install Complete?
No
Empty Pipe Set
MTYmatic
Zero Flow Adjust
Actual Zero
Siemens 2 Channel [1] FLOW1
This menu cell
shows the actual
spacing distance
required between
the transducers.
Use this menu cell to
select the type of
transducer to be
installed.
This menu cell
allows you to select
a mounting mode.
*Use Reflect Mode
whenever possible.
After the transducers are
operating, you can use
this menu cell to
manually set the zero
flow correction.
These menu cells
identify the spacing
indices on the spacer
bar. You must use
these settings to mount
the transducer properly.
This menu cell indicates the
model number of the
appropriate spacer bar for
your transducers and
spacing requirements.
Once you select the
type, use this menu
cell to specify the
transducer size.
Use this menu cell to
inform the meter that you
have completed the
transducer mounting.
After the transducers are
operating, you can use
this menu cell to set the
empty pipe threshold.
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......