6-52
1020NFM-1D
Section 6
To assign functions to Relay 1:
To access the [Relay 1] option list press the <Right Arrow>.
Move the cursor to the desired relay assignment by pressing the <Up/Down Arrow>.
To store selection press <ENTER>.
Assigning Relay 1 Functions
The 1020 provides one alarm relay. Please refer to the appropriate Installation Drawing in Appendix B
for wiring details. Relays respond to any of the alarm conditions/data functions included on the [Relay
1] option list.
6.9.2
RELAY SETUP
Use this menu to assign a function to channel relays. System 1020 supports 2 types of relay outputs,
Alarm Relay and Pulse Relay. Alarm Relay outputs operate in “fail-safe” mode. The relay(s) are ener-
gized under normal conditions - an alarm condition causes the relay(s) to de-energize until alarm
clears. The Pulse Relay output supports totalizer and batch relay functions. The output pulse width is
approximately 200 ms; maximum activation rate is .5 pulse per second. If totalizer pulses exceed this
rate, excess pulses are stored in overflow register. This allows the relay to “catch up” when flow de-
creases enough.
NOTE: The Reset Totalizer [Yes/No] menu command clears all channel totalizers plus the
overflow register described in the last paragraph.
Not Used
Not Active
Power Off
Power Off alarm occurs when power fails
High Flow
Flow rate exceeds high flow setpoint
Low Flow
Flow rate falls below low flow setpoint
Flow Alarm
Flow rate exceeds or falls below flow setpoints
Fault Alarm
System loses receive signal
Soft Fault
Fault condition - memory mode active
Spacing
Transducer spacing needs adjusting
Empty
Empty pipe alarm
Aeration
Aeration percentage exceeds alarm setpoint
Reverse Flow
Flow is in negative direction
BatchTot
Batch/Sample total advances
Interface
Liquid interface setpoint exceeded
Pos Total
Positive total volume advances by 1 digit
Neg Total
Negative total volume advances by 1 digit
Fltwarn
System fault warning
Relay 1 Option List
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......