7-9
1020NFM-1D
Section 7
fx (drive)
29
N (burst length)
5
Ltn in
5.316
Vf max CU
Vs max m/s
Vs min m/s
1355.00
Samples/Cycle
29
Max Damping
Min Damping
HF
>-0.120
Siemens
Dual Path
4SS10G
Site Setup Data
“Automatic” Adjustment Procedure
z
In the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press
the <Right Arrow> and a pop-up [Manual] prompt will appear.
z
Press the <Up or Down Arrow> to select [Automatic] then press <ENTER>.
z
The current measured correction value is displayed (see below).
z
Press <ENTER> again to install this correction value which will now appear next to the [HF]
menu cell.
NOTE: The value shown in the [Automatic] pop-up prompt can not be changed and is for user
information only.
fx (drive)
29
N (burst length)
5
Ltn in
5.316
Vf max CU
Adjustment
Vs max m/s
>Automatic
Vs min m/s
1355.00
Samples/Cycle
29
Max Damping
Min Damping
HF
>-0.000
Siemens
Dual Path
4SS10G
Site Setup Data
fx (drive)
29
N (burst length)
5
Ltn in
5.316
Vf max CU
>Automatic
Vs max m/s
= 0.026
Vs min m/s
1355.00
Samples/Cycle
29
Max Damping
Min Damping
HF
>-0.026
Siemens
Dual Path
4SS10G
Site Setup Data
z
If you decide not to use the [Automatic] selection, press any key other than <ENTER> to abort
the operation.
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......