6-65
1020NFM-1D
Section 6
To enable (or disable) Line Feed:
To access the [Line Feed] option list press the <Right Arrow>.
Press the <Up/Down Arrow> to move the asterisk to [Yes] Line Feed enabled, or [No] Line Feed
disable.
To store the data press <ENTER>. This moves the cursor to [Network ID].
Network ID
The [Network ID] menu cell stores an identification number to facilitate host computer polling when you
use this system in a network environment. The Network ID number can be any value other than 0
(zero). The default setting, zero, disables the network function. Note that entering a non-zero Network
ID number suspends all routine Datalogger activity regardless of any selected options.
NOTE: If you are using the flowmeter for a “stand-alone” application, you must keep the zero
Network ID number.
To assign a Network ID number
To enable numeric entry press the <Right Arrow>.
Use PC numeric keys or KeyPad to enter the [Network ID] number.
To store Network ID press <ENTER>. This moves the cursor to [RTS Key Time].
RTS Key Time
During a serial transmission session, you can select how long the flow computer holds its
request-to-
send
line high until it receives a
clear-to-send
signal. If the RTS key time expires, the 1020 will return its
RTS line low and abort the transmission attempt. Data entry units are
seconds.
The default [RTS Key
Time] is 0.2 seconds.
NOTE: The [RTS Key Time] function is only provided when the meter’s Network ID number is
non-zero.
To set the RTS Key Time:
To access the [RTS Key Time] option list press the <Right Arrow>.
To move the asterisk to the required RTS time press the <Up/Down Arrow>.
To store the data press <ENTER>.
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......