2-18
1020NFM-1D
Section 2
Data Display Mode
After you complete the installation, and the meter is operating normally, you can use <CTRL + L> on
your PC keyboard to toggle between
Installation Menu
mode and
Data
Display
mode (strike <L> while
holding down <CTRL> key). Note that the 1020 RS-232 interface does not support graphics. There-
fore, if you use HyperTerminal to view the data display screens, you will see the data in alphanumeric
form only. You can use the PC’s <Up Arrow> and <Down Arrow> to switch between available display
screens.
2.7.4 DATA ENTRY USING THE RS-232 INTERFACE
The following is a step-by-step example of how to enter data for a Single-Channel 1020N Integral flow
computer using the RS-232 interface and a communications program. Again, don’t let the size of the
Installation Menu intimidate you. Most menu cells contain default parameters. You need only access
the menu cells that control a needed parameter, such as the pipe outer diameter.
Accessing and Leaving the Installation Menu
z
Make sure the Menu Lockout Switch is in the correct position. Connect the serial interface cable
between the flow computer and your laptop computer.
z
Run HyperTerminal (or your choice of communication program) using settings explained previ-
ously.
z
In Command Mode, type [menu 1000]; press <ENTER> to select the top level of the Installation
Menu.
NOTE: NOTE: Type <Menu> to enter Installation Menu. After programming is complete, re-
peated <Left Arrow> key will access the Flow Display. Key <Enter> from Flow Display
to re-enter menu. Key <Ctrl L> to exit menu.
Note on the screen above that there are two columns, a left-hand and a right-hand side. The first left-
hand item, [Meter Type], is highlighted (reversed). [Meter Type] allows you to choose a configuration
from the right-hand column list (on Single Channel models, there is only a single channel)
.
The next
left-hand selection is Meter Facilities. Use this menu to set global options and controls.
Siemens Site [1] Channel 1
Select Meter Type
Meter Type
Single Channel
Meter Facilities
Liquid Class
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......