Upon completion, the meter moves the cursor to [Operation Adjust] on the Installation Menu.
Using The ReversaMatic Routine
As an example, we have successfully installed 1021Transducers, Size C1, in Reflect Mode. However,
conditions at the site will not allow us to change the flow rate at the mounting location. In this case, we
will use the ReversaMatic routine to establish a zero flow level.
NOTE: It is recommended that the Actual Zero routine be performed at the first opportunity. In
addition, You should perform the ReversaMatic procedure as quickly as possible to
ensure that the flow rate remains constant throughout the procedure.
z
Press the <Right Arrow> to access the [Zero Flow Adjust] option list. Move the cursor to [Reversa-
Matic]. Press [ENTER] to invoke the routine.
Siemens 2 Channel [1] SITE1
Transducer Model
1021 Xdcr
Transducer Size
C1
Xdcr Mount Mode
Reflect
Spacing Offset
Invalid
Number Index
4
Spacing Method
Invalid
Ltn Value
2.330
Install Complete?
Yes
Empty Pipe Set
MTYmatic
Zero Flow Adjust
Actual Zero
Integrating 11
Press [ENT]
Pick/Install Xdcr
This counter
increments to sixty
during Zero Flow
sensing.
z
The meter begins to measure the positive flow rate
. “Positive’” flow refers to flow moving from the
upstream transducer location to the downstream transducer location
. Note the top prompt line
shows:
ReversaMatic Active
(see next page).
Siemens 2 Channel [1] SITE1
Transducer Model
1021 Xdcr
Transducer Size
C1
Xdcr Mount Mode
Reflect
Spacing Offset
Nominal
Flow
Number Index
4
Spacing Method
Invalid
Ltn Value (in)
Auto Zero
Install Completed?
ZeroMatic
Empty Pipe Set
ZeroClr
Zero Flow Adjust
Actual Zero
>ReverseMatic
Conforms Indicated flow to Actual zero
Pick/Install Xdcr
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......