7-13
1020NFM-1D
Section 7
T
o clear Active Memory using the KeyPad:
Turn off power (if it is currently on) and the reapply power to the system.
Press and hold ENTER
during the power-on delay. Examine the left-hand figure below and
note the message on the LCD display:
Press
. Note that the greater than (>) symbol changes to a (:) colon. Press to change the
second line to [Yes] (as shown in the right-hand figure above).
Press ENTER
to clear the contents of Active Memory (but preserve any stored Site Setups).
To restore operation, create a new site setup or recall a stored site setup and then re-select any
Meter Facilities items (e.g., RS-232 setup parameters) that you may have edited originally.
7.8
USING THE 1020 RESET SEQUENCE
Should you encounter an operating problem that blocks access to the System Menu, or the meter
operates erratically after exposure to a power transient or some other traumatic event, this might
require the use of the reset sequence in order to restore proper operation. System 1020 provides two
system resets. The first uses the KeyPad and the second method uses a PC command sent via the
RS-232 interface (providing you have access to the flow computer in this way). This section describes
both system reset methods.
The reset sequence operates on two levels. The first level deletes all the data currently in Active Memory,
but leaves Datalogger data and all stored Site Setups intact. This is the most desirable method since
to restore operation all you have to do is reload a saved Site Setup. If this fails then you have to use the
second level, which clears ALL Dynamic Memory.
Be aware that this action erases all saved Site
Setups (including flow calibrated sites), any logged data and any user-defined pipe and transducer
tables.
This will require you to re-initialize the flow computer and transducers and repeat all desired site
settings, custom pipe tables, etc.
7.8.1 SYSTEM RESET USING THE KEYPAD
As mentioned previously, there is a power-on delay of approximately 30 seconds from the time that you
apply power to the 1020 flow computer to when the 16 x 2 Character LCD becomes active. The reason
for the delay is to give you a chance to perform the following system reset procedure.
Wide Beam
Clamp-On
Ultrasonic
Flowmeter
Clr Active Memory? Yes/No
> No
Wide Beam
Clamp-On
Ultrasonic
Flowmeter
Clr Active Memory? Yes/No
: Yes
Wide Beam
Clamp-On
Ultrasonic
Flowmeter
Wide Beam
Clamp-On
Ultrasonic
Flowmeter
Summary of Contents for SITRANS FUS1020
Page 3: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ...21614 C ...
Page 222: ...21614 C ...
Page 223: ...990TNHM 8 OUTLINE DIMENSIONS TRANSDUCERS TRACKS DEDICATED HYBRID MODE 21614 C ...
Page 224: ...1011FTP 8 21614 C ...
Page 225: ...1011FTNF 8 21614 C ...
Page 226: ...21614 C PFA DFT FLOWTUBE INSTALLATION OUTLINE 992DFTP T 8 ...
Page 227: ...INSTALLATION OUTLINE 992 SERIES EXTENDED FLOWTUBE 21614 C 992DFTN 8 ...
Page 228: ......