The I
MPLANT
C
ENTER
2 is prepared for use as follows:
- Make sure that the device is connected correctly
and properly insulated.
- Switch the power switch (Fig. 2-5) to position I
(ON).
- Suspend the bottles or bags of physiological saline
or sterile water from the brackets.
- Raise the irrigation pump covers.
- Insert the irrigation line cassettes horizontally into
their housings.
- Close the covers.
- Insert the irrigation line perforating pins into the
bottles or pouches of physiological saline and open
the cap on the pins.
- Make any necessary setting adjustments (irrigation
flow rate, speed, torque, power, power mode,
program, etc.) (refer to chapter 7).
Note
: This sequence must be completed under
conditions of controlled asepsis for the safety of your
patients. An operating assistant may be needed.
Note
: The data is saved automatically, in order to
maintain the display parameters as previously used
upon switchover from the various modes.
I-S
URGE
L
ED
function
- Check that there are no traces of moisture on the I-
S
URGE
L
ED
micro-motor or micro-motor cord
connections. If there are, remove by wiping and
drying with the multifunction syringe fed with
medical-quality filtered air.
- Connect the micro-motor cord to the lefthand
connector of the device (Fig. 1-7).
- Align the micro-motor and cord connector electrical
contacts, then screw the motor onto the connector.
- Connect a contra-angle to the I-S
URGE
L
ED
micro-
motor.
- Fit an instrument (drill, bur, etc.) to the contra-
angle.
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the contra-
angle.
- Run the motor fitted with its contra-angle and its
instrument without irrigation at moderate speed for
10 to 15 seconds in order to diffuse the lubricant
and eliminate excess.
- Prime the irrigation circuit by pressing the control
keypad flush/prime button (Fig. 3-1).
- When the irrigation solution reaches the end of the
contra-angle, release the footswitch.
- Use I
MPLANT
C
ENTER
2 in accordance with good dental
practice.
P
IEZOTOME
function
- Connect the P
IEZOTOME
2 L
ED
handpiece cord to the
righthand connector of the device (Fig. 1-9).
- Screw the selected tip onto the handpiece using a
tip wrench (according to the recommendations of
the clinical booklet).
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the
P
IEZOTOME
2 L
ED
handpiece.
- Prime the irrigation circuit by pressing the
flush/prime button on the screen (Fig. 3-1) or the
multifunction footswitch (Fig. 11-2).
- When the irrigation solution reaches the end of the
tip, release the footswitch.
- Use the I
MPLANT
C
ENTER
2 in accordance with good
dental practice.
N
EWTRON
function
- Check that there are no traces of moisture on the
N
EWTRON
L
ED
handpiece or the scaler cord
connections. Remove any moisture by wiping and by
drying with the multifunction syringe fed with
medical-quality filtered air.
- Connect the scaler cord to the righthand connector
of the device (Fig. 1-9).
- Align the electrical contacts of the N
EWTRON
L
ED
handpiece and the scaler cord, then plug the
handpiece into the cord.
- Screw the selected tip onto the handpiece using a
tip wrench (according to the
T
IPBOOK
recommendations).
- Fasten the irrigation line to the cord using line
clips.
- Connect the end of the irrigation line to the scaler
cord.
18
Summary of Contents for Implant Center 2
Page 2: ...2 ...
Page 29: ...29 ...
Page 55: ...55 ...
Page 81: ...81 ...
Page 107: ...107 ...
Page 133: ...133 ...
Page 159: ...159 ...
Page 185: ...185 ...
Page 211: ...211 ...
Page 237: ...237 ...
Page 244: ...244 15 16 Fig Afbeelding 5 13 14 Fig Afbeelding 4 ...
Page 245: ...245 20 21 22 23 19 18 24 17 Fig Afbeelding 6 25 26 27 28 29 30 31 Fig Afbeelding 7 ...
Page 246: ...246 41 Fig Afbeelding 9 38 39 40 32 33 34 35 36 37 Fig Afbeelding 8 ...
Page 247: ...247 43 44 45 42 48 49 50 46 47 52 51 Fig Afbeelding 10 1 3 4 5 2 Fig Afbeelding 11 ...