VIII - SAFETY
The I
MPLANT
C
ENTER
2 is equipped with a self-diagnosis
system to detect malfunctions.
When an excessively high temperature of the I-S
URGE
L
ED
micro-motor is detected, symbol 1 (chap. XVII -
Symbol) is displayed in the error zone (Fig. 3-9) and
the device beeps four times.
The device switches to degraded mode so that it can
complete the dental procedure.
The available torque value is set at 25% in order to
protect the I-S
URGE
L
ED
micro-motor.
It is recommended that the I-S
URGE
L
ED
micro-motor
be allowed to cool down until symbol 1 (chap. XVII -
Symbol) disappears.
In the case of a motor malfunction, symbol 2 (chap.
XVII - Symbol) is displayed in the error zone and the
device beeps four times.
It is recommended that you check the connections of
your motor and your motor cord. If the fault persists,
switch the device off at the power switch (Fig. 2-5)
and switch it back on again.
In the case of an internal operation malfunction,
symbol 3 (chap. XVII - Symbol) is displayed in the
error zone and the device beeps four times.
It is recommended that the device be switched off
(Fig. 2-5) and switched back on again.
In the event of a handpiece detection fault, symbol
n°4 lights up in the error zone and four beeps sound.
It is recommended to check:
- The handpiece connection to the connector of the
device.
- The selected mode on the device (left / right)
should be on the same side as the handpiece
connector.
IX - USING THE DEVICE
Important:
- Do not disconnect the micro-motor cord or the
scaler cord when the device is switched on and
the footswitch is pressed.
- Do not disconnect the micro-motor or the
ultrasonic handpieces when the device is
switched on and the footswitch is pressed.
- Do not connect the instrument to, or disengage it
from, the contra-angle when the micro-motor is
running.
- Do not screw or unscrew tips when the
handpieces are on.
- All accessories must be cleaned, disinfected and
sterilized before use.
For your safety and that of your patient, I
MPLANT
C
ENTER
2 must not be used with accessories other
than those supplied or recommended by SATELEC.
- Check the integrity of the device and its
accessories before and after each use in order to
detect any problem.
- If necessary, do not use the device and replace
any faulty component.
17
Program Power
Irrigation
Main functions
Very
powerful
D1
3
60 ml/min.
Osteotomy,
osteoplasty
Powerful
D2
3
60 ml/min.
Medium
D3
3
60 ml/min.
Low
D4
3
60 ml/min.
Soft tissue
detachment
Program
Power
Irrigation
Main functions
Green
Soft
P=5
15 ml/min.
Periodontics
Yellow
Medium
P=5
15 ml/min.
Endodontics
Blue
High
P=5
15 ml/min.
Scaling
Orange
Boost
P=5
15 ml/min.
Loosening
Summary of Contents for Implant Center 2
Page 2: ...2 ...
Page 29: ...29 ...
Page 55: ...55 ...
Page 81: ...81 ...
Page 107: ...107 ...
Page 133: ...133 ...
Page 159: ...159 ...
Page 185: ...185 ...
Page 211: ...211 ...
Page 237: ...237 ...
Page 244: ...244 15 16 Fig Afbeelding 5 13 14 Fig Afbeelding 4 ...
Page 245: ...245 20 21 22 23 19 18 24 17 Fig Afbeelding 6 25 26 27 28 29 30 31 Fig Afbeelding 7 ...
Page 246: ...246 41 Fig Afbeelding 9 38 39 40 32 33 34 35 36 37 Fig Afbeelding 8 ...
Page 247: ...247 43 44 45 42 48 49 50 46 47 52 51 Fig Afbeelding 10 1 3 4 5 2 Fig Afbeelding 11 ...