11. 1 IRRIGATION LINE MAINTENANCE
Sterilizable irrigation line
Refer to the user's manual delivered with the
equipment.
The irrigation line delivered with the I
MPLANT
C
ENTER
2
unit are sterilizable. The perforaters are single-use
and delivered sterile.
Sterile irrigation line
Satelec can provide disposable irrigation lines which
must always be discarded, in safety container for
soiled medical items, after use. Re-use of a single-use
irrigation line may result in contamination of your
patients and expose you to de facto liability. Do not
attempt to resterilize the single-use irrigation lines.
Do not attempt to modify the irrigation lines.
11. 2 CONTRA-ANGLE MAINTENANCE
Refer to the manufacturer's instructions for your
contra-angle.
11. 3 ROTARY INSTRUMENT MAINTENANCE
Please refer to the manufacturer's instructions for
your instruments (burs, drills, etc.).
11. 4 DEVICE MAINTENANCE
Important:
Never use an abrasive cleaning agent on the
device.
Do not use sprays or liquids to clean and disinfect
the I
MPLANT
C
ENTER
2 control unit.
The device, the brackets and the multifunction
footswitch are not sterilizable.
The I
MPLANT
C
ENTER
2 control unit, the control
footswitch and the bracket must always be cleaned
and disinfected after each procedure, using alcohol,
disinfectant agents or disinfectant wipes for dental
practices, such as S
EPTOL
™ wipes.
Regular monitoring of the I
MPLANT
C
ENTER
2 unit is
necessary in order to detect any problem.
It is important to keep the control unit ventilation
vents clean in order to avoid abnormal heating.
11. 5 MICROMOTOR CORD AND SCALER CORD
MAINTENANCE
Important:
Never use an abrasive cleaning agent on the cords.
The cords must not be immersed or disinfected
using agents containing acetone, chlorine or
bleach.
The cords must not be cleaned in an ultrasonic
bath.
a) Cleaning and disinfection
The cords must always be cleaned, disinfected and
sterilized after each procedure, using alcohol,
disinfectant agents or disinfectant wipes for dental
practices, such as S
EPTOL
™ wipes.
b) Sterilization
The cords can be sterilized in an autoclave under the
following conditions:
- Autoclave: Class B
- Sterilization temperature: 134°C at 2 Bar.
- Time at sterilization temperature: 18 minutes.
CAUTION: Not all autoclaves can reach 134°C. Not
all autoclaves draw a pre-vacuum. Please refer to
your autoclave manufacturer for specific
sterilization instructions.
11. 6 I-SURGE LED MICRO-MOTOR MAINTENANCE
Refer to the user's manual delivered with the
equipment.
11. 7 ULTRASONIC HANDPIECE MAINTENANCE
After each use, the handpiece irrigation circuit must
be rinsed with distilled or demineralized water for 20
to 30 seconds.
Disconnect the N
EWTRON
L
ED
handpiece from its cord
before cleaning, disinfection and sterilization.
20
Summary of Contents for Implant Center 2
Page 2: ...2 ...
Page 29: ...29 ...
Page 55: ...55 ...
Page 81: ...81 ...
Page 107: ...107 ...
Page 133: ...133 ...
Page 159: ...159 ...
Page 185: ...185 ...
Page 211: ...211 ...
Page 237: ...237 ...
Page 244: ...244 15 16 Fig Afbeelding 5 13 14 Fig Afbeelding 4 ...
Page 245: ...245 20 21 22 23 19 18 24 17 Fig Afbeelding 6 25 26 27 28 29 30 31 Fig Afbeelding 7 ...
Page 246: ...246 41 Fig Afbeelding 9 38 39 40 32 33 34 35 36 37 Fig Afbeelding 8 ...
Page 247: ...247 43 44 45 42 48 49 50 46 47 52 51 Fig Afbeelding 10 1 3 4 5 2 Fig Afbeelding 11 ...